Studies on Simultaneous Measurement of Pulmonary Membrane Diffusing Capacity (DM), Pulmonary Capillary Blood Volume (Vc) and its Flow (Qc), Using Gas Chromatography:(Report 1) Method and Normal Values Yoshiaki Moriyama 1 1Second Department of Internal Medicine, Niigata University School of Medicine pp.889-897
Published Date 1967/10/15
DOI https://doi.org/10.11477/mf.1404201832
  • Abstract
  • Look Inside

The gas chromatograph modified of Mura-kami's method has been used to analyze gases for simultaneous measurements of pul-monary membrane diffusing capacity (DMco), pulmonary capillary blood volume (Vc) and its flow (Qc) using the breath-holding tech-nique. The two basic columns, silica gel (30 cm in length) and 5 A molecular sieves (290 cm in length) were arranged in series on a dual-column chromatograph with an empty coiled column (8 m in length) to make a balance of both columns. Accurate repro-ducible analyses for C2H2, CO, Ne, CO2, O2 and N2 were obtained from a single 2ml gas sample.


Copyright © 1967, Igaku-Shoin Ltd. All rights reserved.