Laser flare-cell meter : A second prototype Mitsuru Sawa 1 , Yutaka Tsurimaki 1 , Tadahiko Tsuru 1 , Hiroyuki Shimizu 1 1Dept of Ophthalmol, Jichi Med Sch pp.660-664
Published Date 1988/6/15
DOI https://doi.org/10.11477/mf.1410210414

閲覧方法のご案内

この論文を閲覧するためにはログインが必要になります。
パスワードを忘れた場合 新しくユーザー登録する 施設共通IDをご利用の方はこちらから

PPV購入案内

1,320円 (1,200円+税10%) こちらは、660ページから664ページの文献です。
PPV(ペイ・パー・ビュー)とは、論文(記事)単位で購入・閲覧ができるサービスです。
購入には医書.jp本会員登録が必要です。



会員登録完了後に改めて上記「購入サイトへ」を選択してください。
または「購入サイトへ」選択後に「購入ログイン」、「新規会員登録」の順に進んでください。

会員登録について詳しくはをご確認ください。
  • Abstract
  • Look Inside

We developed an instrument for in vivo quantita-tion of protein concentration and cells in the ante-rior chamber. It consists of three main compo-nents : illumination, measuring device and data processing system.

For the illumination, an He-Ne laser beam is used, with the size of 7×10μm at the place of focus. A photon-counting photomultiplier measures the intensity of scattered beam in the aqueous through a sampling window 0.3×0.5 mm in size. A personal computer controls the system and ana-lyzes the detected signals.

The system has dual modes for measurement of protein concentration and the number of cells. In protein concentration mode, the laser beam is set at 13μW in power output to scan vertically for 0.6 mm. It takes 1 second for the measurement. In cell -count mode, the laser beam is set at 78μW to scan two-dimensionally, 0.6×0.3 mm.It takes 1.2 secondsfor the measurement.

We found a good correlation between the thus measured value and the actual one using human plasma (r=0.993, p=0.00). We also confirmed a good correlation regarding the cell counting mode between the measured value and slit-lamp micro-scopic findings in senile subjects.

In normal eyes in young subjects, values for protein concentration measured 0.13ア0.056 (meanアSD/msec) under mydriatics and 0.15+0. 046 without, corresponding to 3.33 and 3.85 mg/dl. The difference was not significant. Eyes in senile subjects showed significantly higher values, 0.33+0. 15 (p<0.05).

The present system promises to be of clinical value in quantitative measurement of protein con-centration and cells in the anterior chamber when further refined.

Rinsho Ganka (Jpn J Clin Ophthalmol) 42(6) : 660-664, 1988