A case of parotid gland surgery in which facial nerve monitoring by nerve integrity monitor(NIM)was unuseful Kanako Yoshimura 1 , Mayu Uneno 1 , Toru Kamitani 1 , Hikaru Nagao 1 , Ken-ichro Toyoda 1 1Department of Otolaryngology, Kyoto City Hospital pp.1117-1119
Published Date 2017/12/20
DOI https://doi.org/10.11477/mf.1411201453
  • Abstract
  • Look Inside
  • Reference

 We presented a parotid gland surgery case in a patient with myotonic dystrophy that a nerve integrity monitor(NIM)was not useful. In the patient with myotonic dystrophy, muscle relaxant action of rocuronium has been known. We thought why NIM didn't work in this case, because the patient have no peripheral nerve palsy and no instrumental malfunction. We should note that in a case with myotonic dystrophy NIM may not be useful.