雑誌文献を検索します。書籍を検索する際には「書籍検索」を選択してください。

検索

書誌情報 詳細検索 by 医中誌

Japanese

Interobserver differences in assessing glaucomatous optic disc using Heidelberg Retina Tomograph Mayumi Okada 1 , Hidetoshi Tsukamoto 1 , Koji Okada 1 , Hiroki Nii 1 , Michiya Takamatsu 1 , Hiromu K. Mishima 1 1Dept of Ophthalmol, Hiroshima Univ Sch of Med pp.313-316
Published Date 2000/3/15
DOI https://doi.org/10.11477/mf.1410906722
  • Abstract
  • Look Inside

We evaluated the possible interobserver differences in assessing the optic disc in 30 glaucomatous eyes recorded by a Heidelberg Retina Tomograph. The records were analyzed by two glaucoma specialists independently regarding disk area, cup area, cup volume, rim volume and maximum cup depth. There were close correlations between the findings by the two observers in all the five parameters (coefficients of correlation : 0.72-1.00) with no significant interobserver differences (p < 0.05) . The findings show that reproducible interpretation of disc margin is possible when performed by trained observers and even when the manipulation is performed manually.


Copyright © 2000, Igaku-Shoin Ltd. All rights reserved.

基本情報

電子版ISSN 1882-1308 印刷版ISSN 0370-5579 医学書院

関連文献

もっと見る

文献を共有