Morphometry of optic nerve head by Heidelberg Retina Tomograph and optical coherence tomography Keiichiro Iwao 1 , Ryo Iwakiri 1 , Kaori Kobayashi 1 , Hiroshi Kobayashi 1 , Satoshi Okinami 1 1Dept of Ophthalmol,Fac of Med,Saga Univ pp.1923-1927
Published Date 2004/10/15
DOI https://doi.org/10.11477/mf.1410100761
  • Abstract
  • Look Inside

 Purpose:To compare optic nerve head measurements obtained by optical coherence tomography(OCT)and scanning laser tomography. Methods:We measured disc area,cup area,cup/disc area ratio,rim area,rim volume using OCT3000(Humphrey)and Heidelberg Retina Tomograph(HRT)in 50 eyes of 50 normal volunteers. Results:Mean disc area,cup area,cup/disc area ratio,rim area and rim volume was 2.00±0.57mm2,0.50±0.46mm2,0.23±0.15,1.54±0.45mm2and 0.43±0.25mm3with HRT and 2.44±0.43mm2,0.60±0.43mm2,0.24±0.15,1.91±0.43mm2and 0.68±0.53mm3with OCT. There were significant differences in the mean values of all parameter without cup/disc area ratio. A significant correlation coefficients was found in all parameters measured by both systems. The mean coefficients of variation(CV)of disc area,cup area,cup/disc area ratio,rim area,rim volume was 2.60%,5.15%,6.05%,4.75%,and 6.68%in HRT and 4.60%,9.25%,9.77%,6.14%,and 9.02%in HRT. Each system showed good reproducibility. A significant difference was found in the CVs of all parameters. Conclusions:The parameters measured by both systems showed good correlation coefficients. The HRT showed better reproducibility than the OCT. Both systems showed high reproducibility.


Copyright © 2004, Igaku-Shoin Ltd. All rights reserved.