Optics of Stereo-Goniophotography Masakichi Mikuni 1 , Kazuo Iwata 1 , Hiroshi Yaoeda 1 , Shigeru Fujii 1 1Department of Ophthalmology, Niigata University, Medical School pp.165-171
Published Date 1969/2/15
DOI https://doi.org/10.11477/mf.1410204011
  • Abstract
  • Look Inside

Geometrical optics referring to the stereo-go-niophotography was mentioned.

1) The grade of an elevation at the angle of the anterior chamber was able to be measured from the following equation.

△=D・T/M・S+D

△ : elevation to be measured.

D : (distance between the images of the two points on one stereoframe) minus (distance between the images of the two points on the other stereoframe).

T : distance between the surface of the go-nioscopic contact lens and the point on the ang-le of the anterior chamber.

M : total magnification.

S : distance between two pencils at the sur-face of the contact lens.


Copyright © 1969, Igaku-Shoin Ltd. All rights reserved.