Japanese

Clinical evaluation of two types of microkeratome for laser in situ keratomileusis(LASIK) Kosaku Tagawa 1 , Kazuo Kawasaki 2 , Tomomi Higashide 3 1Dept of Ophthalmol,Yatsuo General Hosp 2Jinzu Eye Clin 3Dept of Ophthalmol,Clin Neurosci,Div of Neurosci,Kanazawa Univ Grad Sch of Med Sci pp.1515-1517
Published Date 2004/8/15
DOI https://doi.org/10.11477/mf.1410100698
  • Abstract
  • Look Inside

 We performed laser in situ keratomileusis(LASIK)on 155 eyes using either of two types of microkeratome produced by Moria. Manually operated LSK-ONE was used for 76 eyes. Autorotary M2 was used for 79 eyes. Incidence of irregular flap was significantly less with M2 than LSK-ONE. There was no significant difference between the two microkeratomes regarding:free cap,buttonhole,incomplete pass,canthotomy,corneal epithelial defect,bleeding pannus,epithelial ingrowth,or retreatment. Suction time averaged 24.5±19.8 sec for LSK-ONE and 26.1±5.8 sec for M2. While the average had no significant difference,the deviation was greater for LSK-ONE than M2. The findings show that M2may be safer and easier for LASIK than LSK-ONE.


Copyright © 2004, Igaku-Shoin Ltd. All rights reserved.