雑誌文献を検索します。書籍を検索する際には「書籍検索」を選択してください。

検索

書誌情報 詳細検索 by 医中誌

Japanese

Comparison of two microkeratomes for LASIK Takaki Fukunaga 1 , Shunji Tsuboi 1 1Tsuboi Eye Cent pp.911-914
Published Date 2011/6/15
DOI https://doi.org/10.11477/mf.1410103702
  • Abstract
  • Look Inside
  • Reference

Abstract. Purpose:To evaluate the outcome of LASIK using either of two microkeratomes. Cases and Methods:This retrospective study made on 187 eyes that received LASIK during the past 22 months. Corneal flap was made by M2 Single Use 90®(M2)in 95 eyes and by One Use-Plus SBK®(SBK)in 92 eyes. The two groups were evaluated regarding postoperative visual acuity,thickness of corneal flap and residual stromal bed,and higher-order aberrations. Results:The thickness of corneal flap averaged 110±11μm in SBK group and 128±13μm in M2 group. The difference was significant. There was no difference in thickness of residual stromal bed or in higher-order aberrations. Conclusion:Thinner corneal flap was produced using SBK than M2 as microkeratome for LASIK.


Copyright © 2011, Igaku-Shoin Ltd. All rights reserved.