A case report of runaway pacemaker after six years of implantation Mamoru Tago 1 , Yoshiyuki Tamesada 1 , Sugato Nawa 1 , Yoshimasa Senoh 1 , Shigeru Teramoto 1 1Second Department of Surgery, Okayama University School of Medicine pp.459-461
Published Date 1985/3/15
DOI https://doi.org/10.11477/mf.1404204642
  • Abstract
  • Look Inside

A 61-year-old man was treated by pacemaker implantation due to complete A-V block six years ago. Pacemaker had been working well until he complained tachycardia, headache and palpitation in July 1982. When he visited our hospital ECG record showed the pacing rate of 115/min and "runaway pacemaker" was diagnosed. The cause of runaway pacing seemed to be exhaustion of battery power. The generator was exchanged as soon as he was admitted and the postoperative course is good.

Runaway pacemaker is a rare complication after pacemaker implantation but so critical that the promt treatment is necessary.