Japanese

A case of runaway pacemaker Yoshiaki Osamura 1,2 , Michiaki Hiroe 1 , Masahiko Aosaki 1 , Morie Sekiguchi 1 , Koshichiro Hirosawa 1 , Masayoshi Yokoyama 3 , Shigenobu Koyabu 4 1Dept. of Int. Med., The Heart Institute Japan, Tokyo Women's Medical College 3Dept. of Surgery, The Heart Institute Japan, Tokyo Women's Medical College 4Joko, Inc. pp.193-197
Published Date 1980/2/15
DOI https://doi.org/10.11477/mf.1404203525
  • Abstract
  • Look Inside

Runaway pacemakers are an uncommon com-plication of permanent cardiac pacing. A 70-year-old woman with reccurent syncopal attack due to advanced A-V block has treated with AMTI permanent pacemaker (Model DB-401) implantation. 4 days after implantation, she suddenly collapsed. An electrocardiogram demon-strated the pacemaker spike at a variable rate between 125 to 1000 per minute, and 150 to 160 beats per minutes of ventricular activation. Prompt surgical change of the pulse generator was performed and the patient was recovered. Testing of failed pulse generator by the manu-facture revealed battery failure due to leakage of condenser unit.