雑誌文献を検索します。書籍を検索する際には「書籍検索」を選択してください。

検索

書誌情報 詳細検索 by 医中誌

Japanese

Analytical Study of Impedance Cardiogram:Effects of overloading on the left atrium and endotracheal positive pressure on the dz/dt and delta Z waves Tomoko Goto 1 , Yuko Saito 1 , Hidenori Terasaki 1 , Tohru Morioka 1 , Hirotada Katsuya 2 1Dept. of Anesthesiology, Kumamoto Univ. Medical School 2Intensive Care Unit, Kumamoto Univ. Medical School pp.63-68
Published Date 1979/1/15
DOI https://doi.org/10.11477/mf.1404203307
  • Abstract
  • Look Inside

The effect of pulmonary circulation on the wave form of impedance-cardiogram taken from the thoracic cage was studied in dogs. Aortic blood pressure and flow rate, pulmonary arterial pressure and flow rate, left atrial pressure and endotracheal pressure were all recorded on a polygraph, and analyzed according to the various components of the impedance-cardiogram.

As the endotracheal pressure increased, the delta Z wave from the systolic to diastolic phases showed a damped curvature. When the left atrial pressure was increased by inflating the balloon inserted into the atrium, arterial pressure decreased without any change in its flow pattern, and the height of the delta Z wave increased during the diastolic phase more than in the systolic phase. The form of the delta Z wave became biphasic at the beginning and later ex-hibited 3 peaks.


Copyright © 1979, Igaku-Shoin Ltd. All rights reserved.

基本情報

電子版ISSN 1882-1200 印刷版ISSN 0452-3458 医学書院

関連文献

もっと見る

文献を共有