The Accuracy of Portable Instrument for Ultrasound Bladder Capacity Measurement (BVI 5000) Yuji Kato 1 , Youichi Sue 1 , Hiromitsu Fujii 1 , Atsushi Numata 2 , Narumi Taniguchi 2 , Shigeo Kaneko 2 , Sunao Yachiku 2 , Mitsuhiro Mizunaga 3 , Masanobu Miyata 4 Keyword: 残尿 , 超音波検査 pp.459-462
Published Date 2000/5/20
DOI https://doi.org/10.11477/mf.1413902990
  • Abstract
  • Look Inside

We evaluated a portable instrument for ultrasound bladder capacity measurement (Bladder ScanTM BVI 5000) to estimate residual urine volume in 11 male volunteers. There are highly significant linear correlation of measured volume with actual voiding volume (Y=0.691x+39.14, r=0.909, p<0.001). The accuracy of this instrument was clinically acceptable. It is necessary to evaluate the accuracy of this instrument in children and patients in the future.