Vesicourethral Anastomotic Suture Placement Using"Jichi Needle Pusher"" Osamu Muraishi 1 , Chiaki Tanaka 2 , Toshio Yamashita 3 , Shinya Ishikawa 4 , Kazumi Suzuki 4 , Masahiro Yashi 1 , Masayuki Yuzawa 1 , Tatsuo Morita 1 , Akihiko Tokue 1 Keyword: 前立腺摘除術 , 膀胱形成術 pp.127-131
Published Date 1998/2/20
DOI https://doi.org/10.11477/mf.1413902237
  • Abstract
  • Look Inside

A new device"Jichi Needle Pusher"was developed tofacilitate placement of the anastomotic urethral suturesfollowing retropubic radical prostatectomy or cystectomy.A half circle needle attached to the suture material is setinside the curved barrel which is connected to the tip ofthe shaft of Jichi Needle Pusher. A thin stainless steelspring linked with pusher rod pushes out that needle fromthe barrel driving the needle through the urethral lumen.