A SIMPLIFIED RADIOIMMUNOASSAY FOR PLASMA TESTOSTERONE Masafumi Shirai 1 , Shotaro Matsuda 1 , Mamoru Nakamura 2 , Kenzo Yonezawa 3 1Department of Urology, Tohoku University School of Medicine 2Department of Radiology, Tohoku University School of Medicine 3Department of Hospital Pharmacy, Tohoku University School of Medicine pp.269-273
Published Date 1974/3/20
DOI https://doi.org/10.11477/mf.1413201785
  • Abstract
  • Look Inside

The practical use and precision of our modificated radioimmunoassay for measuring plasma testosterone by simplifing the extract method and omitting chromatography, was evaluated. The results were fair with coefficient of variation of 4.8% and average recovery rate of 94.5±2%. Plasma testosterone of normal males ranged between 430 and 1,180ng/dl by this method with the average of 628±236ng/dl. On the other hand, the value obtained by normal females was 36±1.9ng/dl and the average value obtained by water blank was 1.9±1.2ng/dl.