雑誌文献を検索します。書籍を検索する際には「書籍検索」を選択してください。

検索

書誌情報 詳細検索 by 医中誌

Japanese

Reliability indices of Humphrey Field Analyzer for glaucoma Hideko Totsuka 1,2 , Tadashi Hatakeyama 1 1Hatakeyama Eve Clinic 2Dept of Ophthalmol, Sch of Med Shinshu Univ pp.938-942
Published Date 1990/6/15
DOI https://doi.org/10.11477/mf.1410908211
  • Abstract
  • Look Inside

We evaluated the visual field in 104 eyes with glaucoma and in 35 eyes with ocular hypertension. We used the program central 30-2 of Humphrey Field Analyzer 630.

The global indices were consistent with the pro-gression of glaucomatous visual field defects. Reli-ability was compromised more in fixation loss and less in false positive response. Stage of field loss didnot affect the fixation losses or false positive rates. False negative response increased in frequency along with progression of glaucoma. Short-term fluctuations became larger in the presence of glaucomatous field defect. They were positively correlated with the pattern standard deviation.

False negative rates and short-term fluctuations are two major factors that affect the reliability of glaucomatous visual field defects as tested with the current automated perimeter.


Copyright © 1990, Igaku-Shoin Ltd. All rights reserved.

基本情報

電子版ISSN 1882-1308 印刷版ISSN 0370-5579 医学書院

関連文献

もっと見る

文献を共有