Evaluation of glaucomatous disc by computerized digital color analyzer Kazuaki Tetsumoto 1 , Hiroyuki Miyazawa 1 , Kuniyoshi Mizokami 1 , Setsuro Maruyama 2 , Masatsugu Imamichi 2 1Dept of Ophthalmol, Kobe Univ Sch of Med 2Sun Contact Lens Co, Ltd pp.1921-1924
Published Date 1989/12/15
DOI https://doi.org/10.11477/mf.1410211069
  • Abstract
  • Look Inside

We evaluated the glaucomatous pallor of the optic disc by applying a new computerized color analyzer to the photograph of the fundus. We used a personal computer, PC-9801 and our own soft-ware.

Quantitative analysis of the pallor was perfor-med by dividing the disc area into 4 sections. Color of the disc area was expressed as sets of 3 parame-ters for red, green and blue respectively.

We assessed the reproducibility of the method by analyzing one fundus photograph five times as intraphotographic study, and five photographs of one eye as interphotographic one. The coefficient of variation was 2.3% to 3.2% for the fomer and 3.0% to 3.9% for the latter.

We applied the method to eyes with glaucoma. The analyzer led to detection of subtle changes which corresponded to visual field defect by Octo-pus automated perimeter but which could not be identified by ophthalmoscopy. We assume that our computerized color analyzer is of value for quanti-tative assessment of glaucomatous optic disc.


Copyright © 1989, Igaku-Shoin Ltd. All rights reserved.