雑誌文献を検索します。書籍を検索する際には「書籍検索」を選択してください。

検索

書誌情報 詳細検索 by 医中誌

Japanese

Higher-order aberrations in eyes implanted with intraocular lens of different designs Mutsumi Fuchihata 1 , Yoshihiko Ninomiya 1 , Atsuko Takehara 1 , Sanae Shiga 1 , Naoyuki Maeda 2 1Dept of Ophthalmol, Yukioka Hosp 2Dept of Vis Sci, Osaka Univ Grad Sch of Med pp.1255-1259
Published Date 2007/7/15
DOI https://doi.org/10.11477/mf.1410101847
  • Abstract
  • Look Inside
  • Reference

Abstract. Purpose:To compare higher-order aberrations in eyes implanted with intraocular lens(IOL)of different material and designs. Material and Method:This study was made on IOL-implanted 165 eyes. The series comprised 49 eyes implanted with acryl IOL(SA60AT), 29 eyes implanted with acryl IOL(MA30BA), 26 eyes implanted with silicone IOL(ClariFlex), and 51 eyes implanted with aspheric IOL(KS-3Ai). All the eyes were examined by a wave-front analyzer 6 months after surgery. Results:Spherical aberration was significantly smaller in eyes implanted with aspherical than spherical IOLs. Silicone IOLs showed smaller spherical aberrations than acrylic IOLs but no difference in coma aberration and total higher-order aberrations. Conclusion:Measurement of higher-order aberrations in the eye and the cornea may enable more precise evaluation of optical properties of implanted IOLs.


Copyright © 2007, Igaku-Shoin Ltd. All rights reserved.