Purpose:To describe a new acrylic foldable intraocular lens(IOL)for suturing the haptics to the ciliary sulcus in cataract surgery with no zonular support. Design:The diameter of its optics is 6.5 mm with an overall length of 13.0 mm. A point-like depression serves for fixation. A larger depression on the other side is designd to allow room for the suture to slip and to facilitate correction of position of IOL. This IOL was used in 10 porcine eyes. Result:The IOL could be inserted in a folded state through the 4.0 mm corneoscleral wound. It could be fixated in-the-bag or suture-fixated to the ciliary sulcus. Depression in the haptics allowed correction of position of IOL. Conclusion:This IOL promises to be of value in small-incision multipurpose cataract surgery.