A New Analyzer for Defective Cardiac Pacemaker Tsuyoshi Akanuma 1 , Hideo Matsuzawa 1 , Shoji Eguchi 1 1The Second Department of Surgery, Niigata University School of Medicine pp.287-290
Published Date 1981/3/15
DOI https://doi.org/10.11477/mf.1404203737
  • Abstract
  • Look Inside

With a new analyzer, which we made and consists of constant-electric-current-device and synchroscope, voltage and capacitance of the con-denser and resistance in the output circuit of the cardiac pacemaker can be calculated.

A lithium-battery-pacemaker which was re-moved from a girl about one year and half after implant operation was examined with this analizer and the condenser failure parallel with battery was revieled. This puls generator was devided into sections and also examined and the condenser failure was prooved again.


Copyright © 1981, Igaku-Shoin Ltd. All rights reserved.