Evaluation of various tips for phacoemulsification-aspiration Hiroko Shima 1 , Hiroyuki Amano 1 , Takafumi Kan 1 , Tomomi Yamaguchi 1 1Dept of Ophthalmol, Nagahama City Hosp pp.1155-1159
Published Date 1999/6/15
DOI https://doi.org/10.11477/mf.1410906411
  • Abstract
  • Look Inside

We used three different tips in phacoemulsification-aspiration on 295 cataract eyes. They were : Normal 30°, Micro 30° and Mackool 30° . The duration of ultrasound application was longer with Mackool tip than Normal. The postoperative corneal endothelial cell loss was less with Mackool than the other two tips. This finding seemed to be due to the low heat conductivity of polyimide sleeve of Mackool tip to prevent thermal burn, and to the larger working space in the deep anterior chamber preventing damage to the corneal endothelium by scattered nucleus fragments.