Clinical evaluation of an automated noncontact tonometer Toshio Enoki 1 , Harumi Katsushima 2 , Akira Sone 2 1Dept of Ophthalmol, Sch of Med Nation Defence Med Call 2Dept of Ophthalmol, Sapporo Med Univ Sch of Med pp.681-684
Published Date 1998/4/15
DOI https://doi.org/10.11477/mf.1410905834
  • Abstract
  • Look Inside

We evaluated the clinical usefulness of an automated noncontact tonometer (CT-70, Topcon Co, Tokyo) . Tonometery was performed in 378 eyes of 259 outpatients by CT-70 and Goldmann applanation tonometer (GAT). Average of three consecutive measurements in each eye served as the tonometric value by CT-70. There was a high correlation between the values by CT-70 and GAT (p < 0.0001). There was a tendency for the values by CT-70 to be lower than those by GAT. Following regression line was obtained : CT-70=0.823 + 0.911 × GAT. The standard deviation of difference between pairs of measurements by CT-70 and GAT was 2.1 mmHg. Of all CT-70 readings, 97.4% lay within the permissible tolerance area for noncontact tonometers.


Copyright © 1998, Igaku-Shoin Ltd. All rights reserved.