人間ドック929眼の眼圧を自動非接触型眼圧計CT−50(トプコン社)により連続3回測定し,ゴールドマン圧平眼圧計(GAT)の計測値と比較,検討した。GAT平均値14.7±3.4mmHgに対して,CT−50の平均値は15.2±4.0mmHgであった(p<0.01)。CT−50による3回計測値の91%は眼圧変動幅3mmHg以内にあった。GAT値との相関回帰式はCT−50=0.870×GAT+2.40で,相関係数は最高で0.736(平均値)となり,既報の手動非接触型眼圧計との相関係数より低かった。
We measured the intraocular pressure in 929 apparently healthy eyes enrolled in a health-check program. Tonometry with Goldmann applanation tonometer averaged 14.7±3.4 mmHg (x). Tonometery with a new automated noncontact tonometer CT-50 (Topcon, Tokyo) averaged 15.2±4.0mmHg (y). Both methods gave similar readings (p<0.001). In 91% of the eyes, the readings of consecutive three measure-ments by CT-50 in each eye were within±3mmHg. The regression equation was y=0.870x+2.40. The best correlation coefficient between both methods was 0.736. This value was lower than reported values by manually operated noncontact tonometers.