雑誌文献を検索します。書籍を検索する際には「書籍検索」を選択してください。

検索

書誌情報 詳細検索 by 医中誌

Japanese

Longitudinal changes in refraction after implantation of AvanseeTM YP2.2R(NS1) Kaori Momozaki 1 , Ryo Iwakiri 1 , Isao Nakao 2 , Hiroshi Enaida 2 1Department of Ophthalmology, National Hospital Organization Ureshino Medical Center 2Department of Ophthalmology, Saga University Faculty of Medicine pp.1203-1209
Published Date 2019/9/15
DOI https://doi.org/10.11477/mf.1410213267
  • Abstract
  • Look Inside
  • Reference

Abstract Purpose:To report changes in refraction following implantation of single-piece intraocular lens YP2.2R(NS1).

Cases and Method:NS1 was implanted in 56 eyes during cataract surgery. Refraction was objectively measured one week, 1 and 3 months after surgery. Length of ocular axis averaged 23.35±1.0 mm in the 56 eyes. A constant of 118.6 was used in selecting the power of intraocular lens.

Results:There was no significant change in refraction during 3 months after surgery. There was a tendency towards myopia when compared with the target value of refraction. Significant difference was present between the target and actual refraction after surgery(p<0.01). A constant was 118.28 in the present series of NS1

Conclusion:There was no significant change in refraction during 3 months after implantation of NS1. Eyes implanted with NS1 showed a tendency to be more myopic than the target value. A constant is to be determined for each new type of intraocular lens.


Copyright © 2019, Igaku-Shoin Ltd. All rights reserved.