雑誌文献を検索します。書籍を検索する際には「書籍検索」を選択してください。

検索

書誌情報 詳細検索 by 医中誌

Japanese

New vitreous fluorophotometer Akitoshi Yoshida 1 , Akio Hosaka 1 1Department of Ophthalmology, Asahikawa Medical College pp.1195-1199
Published Date 1984/11/15
DOI https://doi.org/10.11477/mf.1410209288
  • Abstract
  • Look Inside

A high-sensitivity vitreous fluorophotometer (VFP) was developed as a research tool based on a modified Haag-Streit slit lamp (Model 360). Fluorescence is detected by an RCA-4516 photo-multiplier and the resulting signal is processed by low-noise preamplifiers and a logarithmic amplifier. Signals are displayed on an HP 7015B X-Y chart recorder as a semilogarithmic recording.

The range of linearity was spread between fluorescein concentration of 7.0 × 10-11 and 8.0 ×10-I7 g/ml.


Copyright © 1984, Igaku-Shoin Ltd. All rights reserved.