Determination of the Power of the Implant lens Yoshihisa Oguchi 1 1Department of Ophthalmology, Medical School of Keio University 2Department of Ophthalmology, Medical Faculty of Erasmus University, Rotterdam, The Netherlands pp.1107-1111
Published Date 1976/9/15
DOI https://doi.org/10.11477/mf.1410205519
  • Abstract
  • Look Inside

We studied 82 cataract patients prior to sur-gery by echography and keratometry. After cataract extraction and artificial lens implanta-tion, the actual state of refraction of the pati-ents was measured and was compared with the predicted value. In about 80% of the cases, the difference lay within minus or plus one diopter. We investigated into the debatable question whether the implant lens should have the em-metropizing, myopizing, iseiconizing or standard power. We found the iseiconizing power to be theoretically the best and to be best tolerated by the patient.


Copyright © 1976, Igaku-Shoin Ltd. All rights reserved.