Distortion of the Photographed Image in Measurements of the Ocular Fundus Masakichi Mikuni 1 , Hiroshi Yaoeda 1 , Tadashi Shimojima 2 , Jihei Nakagawa 2 1Department of Ophthalmology, School of Medicine, Niigata University 2Olympus Oplical Co., LTD pp.557-564
Published Date 1970/4/15
DOI https://doi.org/10.11477/mf.1410204284
  • Abstract
  • Look Inside

Distortions of the eye and the fundus camera as a combined optical system were computed to see the influence on the measurement of the fundus through analysis of fundus photographs.

1. Distortion of the Gullstrand's schematic eye was calculated and the obtained results were as follows : When the external angle of the light projected into the eye was 5°, the distortion at the Gaussian image plane was-0.095% ; when 10°, -0.383% ; 15°, -0.872% and 20°, -1.574 %.

2. Distortion in case of the optical system composed of the schematic eye and the Olympus fundus camera GRC was shown to be as follows :