Evaluation of an acoustic warning device to inform low fluid level in irrigation solution bag for intraocular surgery Eiichiro Sekii 1 , Masahiro Ishii 1 , Atsushi Mekada 1 1Dept of Ophthalmol,Shiga Univ of Med Sci pp.1655-1659
Published Date 2004/9/15
DOI https://doi.org/10.11477/mf.1410100719
  • Abstract
  • Look Inside

 Background:A system exists to visually monitor the fluid level in the bag containing irrigating solution for intraocular surgery. It is marketed as Opeguard(R)Neokit. This system may be distracting as it has to be frequently watched by the surgeon. It is also inaccurate because of the softness of the bag. Method:We developed a device that automatically emits an alarming sound when the remaining fluid content becomes less than 100ml. We tested the accuracy,usefulness and problems. Doctors in training used this device during actual intraocular surgery. Results:An alarming sound was emitted when the remaining fluid became less than 102±2.5ml. Surgeons could concentrate on the surgery as long as they heard no alarming sound. Conclusion:This device is useful for intraocular surgery.


Copyright © 2004, Igaku-Shoin Ltd. All rights reserved.