Component Disassembly of Bipolar Prostheses due to Bearing Insert Wear. Discussion of the Mechanism Nobuki Sugeno 1 , Shin-ichi Kikuchi 1 , Shigeo Aota 1 , Shoji Yabuki 1 1Department of Orthopaedic Surgery, Fukushima Medical University School of Medicine Keyword: bipolar型人工骨頭 , bipolar prosthesis , ポリエチレンの摩耗 , polyethylene wear , コンポーネントの脱転 , component disassembly pp.1113-1118
Published Date 2008/11/25
DOI https://doi.org/10.11477/mf.1408101402
  • Abstract
  • Look Inside
  • Reference

 Component disassembly of the inner head due to bearing insert wear or damage is one of the disadvantages of bipolar prostheses. We report 3 cases of component disassembly of the inner head of bipolar prostheses of the hip joint. The age of the patients at the time of first hemiarthroplasty was 50 years, 56 years, and 33 years, respectively. The interval between the first operation and the time when the component disassembly of the inner head occurred was 15 years, 6 years, and 19 years, respectively. The outer cup was the touch type in 2 cases and the leaflet type in 1 case. The bearing insert was found to be worn out in all 3 cases.