Basic Experiments to Evaluate the Safety of a Fine-caliber Ultrasonically Activated Device Takahiro KINOSHITA 1 , Eiji KANEHIRA 1 , Kenji OMURA 1 , Yoh WATANABE 1 1Department of Surgery Ⅰ, School of Medicine, Kanazawa University Keyword: 超音波凝固切開装置 , 血管凝固 , 熱損傷 pp.473-478
Published Date 1999/10/15
DOI https://doi.org/10.11477/mf.4426900075
  • Abstract
  • Look Inside

 To evaluate the safety of a fine-caliber (5mm) ultrasonically activated device (AutoSonixTM) with a scissor-type probe (UltraShearsTM), animal experiments were conducted. The arteries of pigs, 3-3.5mm in diameter, were coagulated and divided using this device at the power level of 2. The burst pressure of the artery treated by this device ranged from 192 mmHg to the measuring limit (1242 mmHg) .