Analysis of tissue damage caused by ultrasonically activated device Shunsuke HASHIMOTO 1 , Hozumi TATSUOKA 1 , Hisahiro MATSUBARA 2 , Tadashi YAMAGUCHI 3 , Hideki HAYASHI 3 1Department of Medical System Engineering, Faculty of Engineering, Chiba University 2Department of Frontier Surgery, Graduate School of Medicine, Chiba University 3Research Center for Frontier Medical Engineering, Chiba University Keyword: 超音波凝固切開装置 , キャビテーション , 組織傷害 pp.175-181
Published Date 2010/4/15
DOI https://doi.org/10.11477/mf.4426100461
  • Abstract
  • Look Inside
  • Reference

 In order to determine the safe usage of ultrasonically activated devices(USAD), extent of spreading the biologically harmful“cavitation”from USAD was measured by using the porcine gut. Small intestine was pulled out from 3-month-old domestic pigs under general anesthesia. USAD was activated under several settings, varying the distances between the serosa of the intestine and the tip of the blade. No histological damages were observed on the specimens where the distance between the intestinal serosa and the blade was 0.3 mm or more. However, damages to subserosa and musclaris propria were observed for those specimens with the distance of 0.1 mm and in contact with the serosa. These data suggested that no harmful effect would be expected as long as the blade is apparently apart from the target organs.


Copyright © 2010, JAPAN SOCIETY FOR ENDOSCOPIC SURGERY All rights reserved.