雑誌文献を検索します。書籍を検索する際には「書籍検索」を選択してください。

検索

書誌情報 詳細検索 by 医中誌

Japanese

The effects of variations in the plunger assembly weight upon the tonometric pressure and volume of corneal indentation M. Kageyama 1 , S. Kawamoto 1 1Tokyo Keisatsu Hospital pp.884-887
Published Date 1956/5/15
DOI https://doi.org/10.11477/mf.1410205730
  • Abstract
  • Look Inside

The influence of Schetz's tonometer of which plunger assembly weight including plunger, load, hammer and pointer was out of gauze, against the measuring of intraocular pressure, was indused by the calculation of experimental results of Friedenwald published in decenia reports.

Intraocular pressure and volume of corneal indentation were calculated by the tonome-ters from 5.0gr. P. A. W. to 12.0gr. at 0.5gr. gradation and the numerical value were indi-cated from reading O to 20 of 16 kinds including 15.0gr. P. A. W.


Copyright © 1956, Igaku-Shoin Ltd. All rights reserved.