雑誌文献を検索します。書籍を検索する際には「書籍検索」を選択してください。

検索

書誌情報 詳細検索 by 医中誌

Japanese

Clinical Application of Pulsed Xenon Photocoagulator Kiyonori Amano 1 , Hirohazu Tanaka 1 , Ken-ichi Baba 1 , Shinya Komatsu 1 , Kimiharu Noyori 1 1Department of Ophthalmology, Saitama Medical School pp.823-829
Published Date 1975/7/15
DOI https://doi.org/10.11477/mf.1410205326
  • Abstract
  • Look Inside

A prototype photocoagulator utilizing short arc xenon lamp (500W) and high efficiency aspheric mirror system was constructed. Coagulation time can be set from 0.1 to 0.8 second and the field image at 2, 3, 4, 5, 6 degrees. The output of the xenon lamp can be overcharged up to 1,450 W. Clinically treated cases totalled 22 eyes, including diabetic retinopathy, occlusion of retinal vein, peripheral retinal tear, macular hole and central serous retinopathy. The stan-dard setting was 3 degrees/940 W/0.5 second for posterior lesion and 2 degrees/1,450W/0.3 second for macular lesions.


Copyright © 1975, Igaku-Shoin Ltd. All rights reserved.

基本情報

電子版ISSN 1882-1308 印刷版ISSN 0370-5579 医学書院

関連文献

もっと見る

文献を共有