Prototype of ophthalmoscope with built-in optotypes and filters for examination of fixation Hiroshi Takahashi 1 , Masako Yoshida 2 , Akio Tabuchi 3 , Atsushi Tanabe 4 1Dept of Ophthamol, Kitakyushu Rehab Cent for Children with Disabilities 2Yoshida Eye Clin 3Dept of Sensory Sci, Fac of Health Sci and Technol, Kawasaki Univ of Med Welfare 4Neitz Instruments Co., Ltd. pp.551-556
Published Date 2013/4/15
DOI https://doi.org/10.11477/mf.1410104681
  • Abstract
  • Look Inside
  • Reference

Abstract. Purpose:To describe a modified ophthalmoscope to examine the status of fixation. Cases and Method:A commercially available ophthalmoscope with built-in optotypes were fitted with yellow, red and absorptive filters. Glare and readability of optotypes were judged by 12 patients with visual problems. Five had retinitis pigmentosa, 5 had cone dystrophy, one had choroidal neovascularization, and one had Leber optic neuropathy. Central scotoma was present in 7, concentric visual field in 4, and ring scotoma in one. Results:All the cases complained of glare when no filter was inserted. Use of yellow filter resulted in glare and poor recognition of optotypes. Use of red filter showed less glare and better recognition. Use of absorptive filter showed further reduction of glare and improved recognition. Conclusion:This modified ophthalmoscope promises to be of value in the examination of status of fixation.


Copyright © 2013, Igaku-Shoin Ltd. All rights reserved.