Clinical evaluation of IOL Master® Version 5 Asami Toki 1 , Toshihiro Matsumoto 1 , Mari Yoshikawa 1 , Mayumi Sato 1 , Miyuki Eda 1 , Aya Nozoe 1 1Dept of Ophthalmol,Chigasaki Cent Hosp pp.79-84
Published Date 2010/1/15
DOI https://doi.org/10.11477/mf.1410103064
  • Abstract
  • Look Inside
  • Reference

Abstract. Purpose:To evaluate the usefulness of IOL Master® Version 5 in predicting the power of intraocular lens by comparing with A mode ultrasound measurement. Cases and Method:This retrospective study was made on 177 eyes of 132 patients who received cataract surgery by us during a 2-month period. The axial length was measured by IOL Master® and A-mode ultrasound before surgery. Difference of predicted and actual refraction was defined as error. Results:Axial length could not be measured in 22 eyes(12.4%)by IOL Master®. Error of refraction was within ±0.5 diopters in 66% of eyes by IOL Master® and in 60% by ultrasound. The difference was not significant. Conclusion:IOL Master® Version 5 has similar predictive value as A-mode ultrasound regarding the power of intraocular lens in cataract surgery.