雑誌文献を検索します。書籍を検索する際には「書籍検索」を選択してください。

検索

書誌情報 詳細検索 by 医中誌

Japanese

Evaluation of glistening in an acrylic intraocular lens removed six years after insertion Yoko Kuretake 1 , Toshihiro Matsumoto 2 , Mari Yoshikawa 2 , Mariko Shigeto 2 , Mayumi Sato 2 1Dept of Ophthalmol,Yokosuka Hokubu Kyosai Hosp 2Dept of Ophthalmol,Chigasaki Cent Hosp pp.559-563
Published Date 2006/4/15
DOI https://doi.org/10.11477/mf.1410100422
  • Abstract
  • Look Inside
  • Reference

Purpose:Experimental production of glistening particles in acrylic intraocular lens(IOL)removed 6 years after insertion. Case and Methods:A 86-year-old male underwent bilateral cataract surgery with insertion of acrylic IOL. The IOL had to be removed 6 years later due to decentration. The removed IOL showed 100glistening particles per mm2. We placed the explanted and a new acrylic IOL in saline solution. We then observed the two IOLs by shifting the temperature of the solution from 33 to 27 degrees. Finding:The explanted IOL showed glistening immediately after shifting the temperature. The new IOL showed no glistening throughout. Conclusion:Implanted acrylic IOL may show repeated appearance and disappearance of glistening as long as it stays in the eye. After a while,the IOL may undergo structural changes that facilitate appearance of glistening particles.