Non-invasive Measurement of Cardiac Output by the CO2 Rebreathing Method and its Reliability in Clinical Practice Naomitsu Kuji 1 , Osamu Tochikubo 1 , Makoto Shimizu 1 , Kousuke Minamizawa 1 , Kazuo Kimura 1 , Touyou Nihei 1 , Masao Ishii 1 1The Second Department of Internal Medicine, Yokohama City University School of Medicine Keyword: 心拍出量(cardiac output) , 運動負荷試験(exercise testing) , CO2再呼吸法(carbon dioxide rebreathing method) pp.579-585
Published Date 1991/6/15
DOI https://doi.org/10.11477/mf.1404900295
  • Abstract
  • Look Inside

A Commercially available, non-invasive system for estimation of cardiac output by the CO2-rebreathing method (Sensormedicus MMC4400) was evaluated to determine its reliability in clinical practice. Values of cardiac output were obtained at rest and during mild to moderate bicycle ergometer work in patients with is-chemic heart disease or hypertension. Cardiac output measured by the CO2-rebreathing method was significantly correlated with that measured simultane-ously by dye dilution or thermodilution methods.


Copyright © 1991, Igaku-Shoin Ltd. All rights reserved.