Clinical evaluation of the sensor for body activity responsive pacemaker Toru Tanaka 1 , Tomiyoshi Saito 1 , Jiro Suwa 1 , Miyako Kusumoto 1 , Tsukasa Kawamura 1 , Shigeki Ishihara 1 , Yuki Shinomiya 1 , Yuji Miyazawa 1 , Yukio Komatsu 1 1Cardoivascular Center of Sendai pp.433-438
Published Date 1986/4/15
DOI https://doi.org/10.11477/mf.1404204857
  • Abstract
  • Look Inside

Clinical evaluation of the sensor for body activity pacemaker was done in 28-year-old normal man and 79-year-old female (to whom the pacemaker were externally attached on the chest surface) equipped with Holter monitors recorded both intrinsic ECG and pacemaker impulses. There were good correlation bet-ween intrinsic heart rate and pacemaker rate on the movement of upper and lower extremities. However, the sensor had oversensed to the movement of upper extremity attached with pacemaker. The pacemaker has also good sensitivity to ordinary activities in both young and old ones.

Two patients (age 53 and 83) with sick sinus synd-rome and atrial fibrillation with slow ventricular res-ponse underwent permanent implantation. They demon-strated physiological rate responses with daily activities and their general conditions were improved.

An activity responsive pacemaker may be useful in selected patients with atrial fibrillation with slow ventri-cular response and sick sinus syndrome.