A prototype of ophthalmodynamometer Norikiyo Nishikawa 1 , Masakatsu Fukuda 2 , Katsushirou Hoshikawa 3 1Dept of Ophthalmol, Osaka Police Hosp 2Dept of Ophthalmol, NTT West Japan Osaka Hosp 3New Vision Corporation pp.1247-1249
Published Date 2001/6/15
DOI https://doi.org/10.11477/mf.1410909036
  • Abstract
  • Look Inside

We devised a new suction-type ophthalmodynamometer. It incorporated remote-control switches, fixation-device of suction cup onto the sclera, and two warning signals when the suction pressure reached 300mmHg and 550mmHg levels. A pressure-release switch was installed to enhance the safety of the system. It was easy to operate. The measured central retinal artery pressure was consistent and reliable. The present device promises to be clinically useful.