A prototype of apparatus for objective assessment of stereoscopic vision Keiko Nosue 1 , Yasuko Saito 1 , Tadao Miyashita 1 1Dept of Ophthalmol Toho Univ Sch of Med pp.371-374
Published Date 1990/3/15
DOI https://doi.org/10.11477/mf.1410908112
  • Abstract
  • Look Inside

We designed and built a prototype of apparatus for objective measurement of stereopsis by incor-porating two visual display terminals and a half mirror. Stereoscopic vision was measured in 6 healthy adults by changing the visual acuity in one 'eye with an occluder. There was a decrease in the rate of stereopsis along with a decrease in the visual acuity of non dominant eye. We applied themethod to a group of 33 patients with strabismus or amblyopia and compared the findings obtained by synoptophore. Both methods showed similar results in 29 patients, while our method only was successful in 4 patients. our apparatus seemed to be of particu-lar value in the evaluation of stereopsis in infants.


Copyright © 1990, Igaku-Shoin Ltd. All rights reserved.