Basic study on enzyme-linked immunosorbent assay for adenoviral conjunctivitis Xiaohong Xia 1 , Koki Aoki 2 , Tsuzuki WangLiping 1 , Yutaka Isobe 1 , Xia Xiao 1 , Hanayo Ikue 3 , Yoshiyuki Mochida 3 , Takayuki Kamikochi 4 , Shigeaki Ohno 1 1Dept of Ophthalmol, Yokohama City Univ Sch of Med 2Aoki Eye Clinic 3Special Reference Laboratories 4Toray Fuji Bionics pp.239-243
Published Date 1995/2/15
DOI https://doi.org/10.11477/mf.1410904171
  • Abstract
  • Look Inside

Adenoclone® has been widely used as a rapid and simple kit for detecting adenovirus antigens. However, its positive rate is rather low comparing with that of virus culture. Therefore, the sensitivity of Adenoclone®, which is related to reaction times and serological virus types, was investigated. We also analyzed whether there is any difference of sensitivity in different lot numbers. Adenoclone® was shown to be more sensitive to adenoviruses type 3 and 8 than adenoviruses type 4, 11, 19 and 37. As for the two-step reaction times, the optical density of 3 hours combined with 30 minutes was about 2 times as high as those of 1 hour combined with 10 minutes. No difference of sensitivity related to lot difference of Adenoclone(R)was discovered in our clinical examination.


Copyright © 1995, Igaku-Shoin Ltd. All rights reserved.