雑誌文献を検索します。書籍を検索する際には「書籍検索」を選択してください。

検索

書誌情報 詳細検索 by 医中誌

Japanese

A new optical system for corneal incision with excimer laser Yoichi Manabe 1 , Hideyo Kandori 1 , Tatsuo Yamaguchi 2 , Kiyonori Amano 1 , Kimiharu Noyori 1 1Dept of Ophthalmol Saitama Med Sch 2Dept of Ophthalmol St. Lukes International hosp pp.1699-1702
Published Date 1989/10/15
DOI https://doi.org/10.11477/mf.1410211025
  • Abstract
  • Look Inside

We developed a new optical system for corneal ablation with excimer laser at 193 nm wavelength. An auxiliary helium-neon laser is used for aiming. The system offers a choice of 6 shapes of corneal incisions including linear, circular, oral and triangle ones.

We applied the system to produce linear corneal incisions in the rabbit. The ablation time was reduced to one half because of the new system.Histological examinations showed the surface of the wound to be more smooth and wider than when incised with the non-optical system. With the pres-ent system, the corss section of the wound was V -shaped. The depth of incision was deeper in the central than in the peripheral cornea.

Movement of the eye during treatment seemed to be a major factor for depth of the incision. Our new system seemed to be of value in shortening the time of ablation and in allowing to achieve more consis-tent effects.


Copyright © 1989, Igaku-Shoin Ltd. All rights reserved.

基本情報

電子版ISSN 1882-1308 印刷版ISSN 0370-5579 医学書院

関連文献

もっと見る

文献を共有