  • Look Inside

We examined the finishing of posterior chamber lenses with the use of specular microscopy. Products by 9 manufacturers were tested. We could detect surface contamination, wavy reflex suggesting ir-regular surface, polishing marks, sharp lens edges, and irregular edges of the positioning holes in themajority of the lenses.

Our findings indicate that the finishing of the currently available intraocular lenses are not per-fect when thus examined. Ophthalmic surgeons are warned over the present, less-than-ideal standard of lens finish.

Rinsho Ganka (Jpn J Clin Ophthalmol) 41(6) : 699-702, 1987