Test results of OSCAR color vision tester Takahiko Motohashi 1 , Shinobu Noyori 1 , Kaneo Shimizu 1 , Yasuo Oota 1 , Akira Seki 1 , Tadashi Miyamoto 1 1Dept. of Ophthalmol., Tokyo Med. Coll. pp.1203-1206
Published Date 1985/10/15
DOI https://doi.org/10.11477/mf.1410209547
  • Abstract
  • Look Inside

We examined 83 subjects with congenital color vision defects using OSCAR (objective screeningcolor anomalies and reduction) color vision tester developed by Estevez (1983). The findings were compared with those by anomaloscope and pseudo-isochromatic color plates, Examinations by OSCAR allowed the classification of color vision defects into protan and deutans at a fair degree of accuracy comparable with anomaloscope. It gave poor results in estimating the degree of color vision defect.Because of the ease of operation, the device could be used by infants of 4 years or more in our pre-sent series. These features seem to be excellent fe-atures of the OSCAR system.


Copyright © 1985, Igaku-Shoin Ltd. All rights reserved.