Simultaneous measurement of checkerboard pattern reversal electroretinogram and visually evoked response under direct observation by using infrared television fundus camera Seiichi Totsuka 1 , Tsztyoshi Tanino 1 , Kazuyuki Kurihara 1 , Katsuzo Segawa 1 1Department of Ophthalmology, School of Medicine Shinshu University pp.731-735
Published Date 1983/6/15
DOI https://doi.org/10.11477/mf.1410208928
  • Abstract
  • Look Inside

A checkerboard pattern reversal stimulator was built in an infrared fundus camera. Checkerboard sinusoidal pattern reversal stimulus was elicited by a polaroid sheet. The stimulus frequency, size and intensity was alterable. This pattern stimulus was led into the optical pathway of the infrared televi-sion fundus camera and was projected onto the sub-ject's fundus. The state of fixation and the fundus view was observed by infrared illumination without influencing the state of adaptation of the retina. Response from the eye of normal subject by sclero-corneal contact lens electrode showed a sinusoidal waveform.


Copyright © 1983, Igaku-Shoin Ltd. All rights reserved.