雑誌文献を検索します。書籍を検索する際には「書籍検索」を選択してください。

検索

書誌情報 詳細検索 by 医中誌

Japanese

An Automatic Objective Refractive Dioptron TM Yoshimi Sugimachi 1 , Hisae Matsuoka 1 , Atsushi Kanai 1 , Kazuo Kato 1 , Akira Nakajima 1 1Department of Ophthalmology, Juntendo University, School of Medicine pp.1203-1207
Published Date 1976/10/15
DOI https://doi.org/10.11477/mf.1410205532
  • Abstract
  • Look Inside

Dioptron TM (an automatic objective refractor, Coherent Radiation, U. S. A. Cal.) has been evaluated for its reliability, accuracy and prac-tical usefulness. It has been tested at two eye clinics (on 107 patients) , a primary and a middle school (on 43 students, ranging 6 to 15 years of age) and on 597 policemen without refractive error.

It has measured over 75% of adult subjects or patients with "good reliability". In 6-years-old children, however, over 50% of the cases failed measurements because of "poor relia-bility" or "rejection", although the number decreased with the increase in age.