Newly developed dark adaptometer using a pair of goggles Yukitaka Uji 1 1Nat Inst of Sens Org, Nat Tokyo Med Cent pp.97-101
Published Date 2013/1/15
DOI https://doi.org/10.11477/mf.1410104576
  • Abstract
  • Look Inside
  • Reference

Abstract. Purpose:To report the design and function of newly developed dark adaptometer using a pair of goggles. Method:The present system has a white light emitting diode(LED)built in the goggles to serve as stimulus for prior light adaptation and during dark adaptation. The patient was asked to respond by pushing a button when the light stimulus is noticed during the course of adaptometry. The response is displayed as dark adaptometric curve on a liquid crystal touch panel. The response by a 57-year-old normal person was used to set the baseline conditions of the system regarding the Kohlrausch point and first and second dark adaptation curve. Ten normal persons, aged from 27 to 36 years, were tested under the same condition. Results:Responses by the 10 normal persons closely simulated the standard dark adaptation curve. Conclusion:The findings show the usefulness of the present system as dark adaptometer.