Merits and problems of VISULUXTM during buckling surgery for retinal detachment Takashi Koshino 1 , Shigeru Hoshiai 1 , Shigeki Suzuki 1 , Tadashi Orihara 1 , Narihisa Hanada 1 , Hideyuki Kurihara 1 1Kurihara Eye Hosp pp.965-968
Published Date 2007/6/15
DOI https://doi.org/10.11477/mf.1410101790
  • Abstract
  • Look Inside
  • Reference

Abstract. Purpose:To report outcome of surgery for retinal detachment using VISULUXTM adjunct to operating microscope. Cases and Method:This study was made on 45 eyes that received scleral buckling for rhegmatogenous retinal detachment during the past 2 years. Operating microscope with VISULUXTM was used in all the eyes. Findings were compared with another series of 34 eyes that received retinal detachment surgery under indirect ophthalmoscope. Results:Initial reattachment was obtained in 41 out of 45 eyes(91%)and 30 out of 34 eyes(88%)respectively. No significant difference was present between the two groups. Duration of surgery was similar in the two groups. Intraoperative visibility of the fundus was compatible with that of slit-lamp microscope. Conclusion:Use of VISULUXTM as the illuminating system in surgery for retinal detachment ensures good intraoperative visibility.


Copyright © 2007, Igaku-Shoin Ltd. All rights reserved.