Design of a Targeting Device for Low-Intensity Pulsed Ultrasound (LIPUS) Therapy. Creation of a New Targeting Device Akira IKUMI 1 , Harumitsu ICHIMURA 1 , Seira YAMAMOTO 1 , Eriko OKANO 1 , Norihito ARAI 1 , Masafumi UESUGI 1 , Ikuo AITA 1 1Department of Orthopaedic Surgery, Tsukuba Medical Center Hospital Keyword: 低出力超音波パルス , low-intensity pulsed ultrasound , ターゲットデバイス , targeting device , 単純X線撮影 , plain radiography pp.753-757
Published Date 2013/8/25
DOI https://doi.org/10.11477/mf.1408102785
  • Abstract
  • Look Inside
  • Reference

 Background:There have been many reports on the shortening effect of LIPUS on the fracture union period, but there have been occasional reports of cases of ineffectiveness in regard to the treatment of humeral and femoral fractures. Some of the cases of ineffectiveness are thought to be due to the targeted fragmented bone parts being farther from the surface of the skin, and therefore not properly irradiated by the ultrasound beam.

 Method:To address this problem we devised a targeting device that makes it easy to aim the irradiation at a specific location by using plain radiography, and we assessed its usefulness.

 Outcome:Targeting with the device we created made it easy to adjust the irradiated position, and it required little time and made it easy to visually confirm the irradiated position for the radiograph.

 Conclusion:We therefore concluded that it is useful for LIPUS therapy, which requires accuracy in terms of the irradiated position and angle.