雑誌文献を検索します。書籍を検索する際には「書籍検索」を選択してください。

検索

書誌情報 詳細検索 by 医中誌

Japanese

Quality Control of Medical Instruments and Cleaning Water Hirobumi NAKAYAMA 1 , Kimiyuki KAWAGUCHI 1 , Akihito KATO 2 , Hiroyuki MORISHIMA 1 , Goro NAGASHIMA 1 , Setsuko YAGIOKA 3 , Takeshi ASAKURA 3 , Shinichi TANAKA 4 , Yuichiro TANAKA 5 1Department of Neurosurgery, St. Marianna University Kawasaki Municipal Tama Hospital, St. Marianna University School of Medicine 2Medical Emergency and Disaster Center, St. Marianna University Kawasaki Municipal Tama Hospital, St. Marianna University School of Medicine 3Operating Room, St. Marianna University Kawasaki Municipal Tama Hospital, St. Marianna University School of Medicine 4Infection Control Team, St. Marianna University School of Medicine 5Department of Neurosurgery, St. Marianna University School of Medicine Keyword: surgical instrument , corrosion , reversed-osmosis(RO)water pp.845-850
Published Date 2019/8/10
DOI https://doi.org/10.11477/mf.1436204035

閲覧方法のご案内

この論文を閲覧するためにはログインが必要になります。
パスワードを忘れた場合 新しくユーザー登録する 施設共通IDをご利用の方はこちらから

PPV購入案内

1,320円 (1,200円+税10%) こちらは、845ページから850ページの文献です。
PPV(ペイ・パー・ビュー)とは、論文(記事)単位で購入・閲覧ができるサービスです。
購入には医書.jp本会員登録が必要です。



会員登録完了後に改めて上記「購入サイトへ」を選択してください。
または「購入サイトへ」選択後に「購入ログイン」、「新規会員登録」の順に進んでください。

会員登録について詳しくはをご確認ください。
  • Abstract
  • Look Inside
  • Reference

 PURPOSE:Herein we aimed to investigate the degradation of surgical instruments in our hospital and how water quality affects the rate of metal corrosion.

 MATERIALS and METHODS:We observed 279 stainless steel instruments, and determined the presence of damage like metal corrosion or scale formation. We also measured the concentrations of chloride(Cl)and silicate(SiO44−)ions in the water used for cleaning in our operating rooms, including tap water from the city water supply and reverse-osmosis(RO)filtered water.

 RESULT:Pitting corrosion was observed on 71% of the instruments we investigated. The concentration of Cl was 0.7mg/L in tap water and 0.1mg/L in RO water, while the concentration of SiO44− was 0.3mg/L in both the tap and RO water.

 DISCUSSION:Of the dissolved ions Cl and SiO44−, Cl is more of a concern, as it causes pitting corrosion over time, while SiO44− causes scale formation. Considering the typical water quality in the operating-room environment, degradation must be monitored for the general maintenance of metal surgical instruments.


Copyright © 2019, Igaku-Shoin Ltd. All rights reserved.