Extracorporeal Ventriculoatrial Shunt with the Use of One-way Ball Valve Shigeki KUBO 1 , Hiroshi TAKIMOTO 1 , Kazuki HOSOI 1 , Shingo TOYOTA 1 , Shuji TAKAKURA 1 , Yasuhiro HAYASHI 1 , Masato UENO 1 , Toshitaka MORISAKO 1 , Jun KARASAWA 1 , Hideo NINAGA 2 , Toshiki YOSHIMINE 3 1Department of Neurosurgery, Osaka Neurological Institute 2Department of Anesthesiology, Osaka Neurological Institute 3Department of Neurosurgery, Osaka University Graduate School of Medicine Keyword: hydrocephalus , ventriculardrainage , ventriculoatrial shunt , extracorporeal , one-way ball valve pp.405-409
Published Date 2002/4/10
DOI https://doi.org/10.11477/mf.1436902198
  • Abstract
  • Look Inside

We developed a simple system of an “extracorporeal” ventriculoatrial (VA) shunt using a one-way ball valve (Acty valve ⅡR, Kaneka Medix) to release the patient from postoperative constraint during the ventricular drainage. The system is constructed in such a way that the ventricular drainage tube is connected to the central venous catheter via a one-way valve. The CSF is regulated by using the valve and is diverted into the systemic circulation as in the conventional ventriculoatrial shunt.